Intrathecal drug therapy using the Codman Model 3000 Constant Flow Implantable Infusion Pumps

H
ETHANS KD; SCHRYVERS OI; NANCE PW
SPINAL CORD , 2005, vol. 43, n° 4, p. 214-218
Doc n°: 118045
Localisation : Centre de Réadaptation de Lay St Christophe , en ligne
Descripteurs : AD32 - SPASTICITE Url : http://www.nature.com/sc/archive/index.html

The objective of this study was to evaluate the accuracy, reliability, safety, and efficacy of the Codman Model 3000 Constant Flow Implantable Infusion Pump for intrathecal baclofen delivery as a therapeutic option for the treatment of severe spasticity. The distinctive features of this pump include a raised, easily palpable septum, a safety valve protecting the bolus pathway, no programmer needed, and no battery to fail. Design: A total of 17 patients with spinal cord injury, multiple sclerosis, or cerebral palsy were implanted with this pump. The accuracy of the pump and drug treatment efficacy was determined at each visit and adjustments to the dosages were made as required. All the intrathecal drug delivery system complications were reviewed. Results: The expected efficacy was achieved. The accuracy of the implanted pumps ranged from 90-97% (average 94%). There were no complications due to primary pump problems. The complications reported are similar to other implantable infusion devices and include dehiscence of the suture line, pressure ulcer development, formation of seroma, inversion of the pump, baclofen overdose, and catheter failures. Conclusion: The Codman Model 3000 Constant Flow Implantable Infusion Pump is an accurate, reliable, and convenient option for patients needing intrathecal baclofen therapy, with complications similar to other available pumps.

Langue : ANGLAIS

Mes paniers

4

Gerer mes paniers

0