Mechanical characteristics of microswitches adapted for the physically disabled

WEISS PL
J BIOMED ENG , 1990, vol. 12, n° SEPT., p. 398-402
Doc n°: 51398
Localisation : Documentation IRR
Descripteurs : KF6 - FAUTEUIL ROULANT

The object of this study to measure the mechanical characteristics of several commonly used adapted switches in order to demonstrate the feasibility of collecting quantitative descriptions of switch performance. A stepping motor was used to drive a digital micrometer vertically towards the tested switch and proving-ring strain gauges recorded the operating force. Measures of activation and deactivation forces, travel and switch compliance were compared several clinical examples illustrate the functional application of these results. It is anticipated that this information will permit clinicians to prescribe switches in a more accurate manner.

Langue : ANGLAIS

Tiré à part : OUI

Identifiant basis : 1991000430