Intra and intersession repeatability and reliability of the S-Plate(R) pressure platform

IZQUIERDO RENAU M; PEREZ SORIANO P; RIBAS GARCIA V; QUERALT A
GAIT POSTURE , 2017, vol. 52, p. 224-226
Doc n°: 183587
Localisation : Documentation IRR

D.O.I. : http://dx.doi.org/DOI:10.1016/j.gaitpost.2016.12.001
Descripteurs : DF21 - GENERALITES - MARCHE

The aim of this study was to assess the repeatability and reliability of the
S-Plate(R) pressure platform in a group of healthy subjects.
METHODS: Forty subjects, free from physical conditions that would affect normal
gait, walked along a five-meter corridor while data were recorded from the
pressure platform. A total of 10 steps (five each side) were obtained as well as
five static trials; the same measurements were repeated one week later. Peak and
mean plantar pressures and contact area were recorded for both dynamic and static
trials. Additionally, weight supported on each limb was documented during static
trials. To assess intrasession and intersession repeatability and reliability,
the intraclass correlation coefficients (ICCs) and coefficient of variation (CoV)
were calculated. RESULTS: Taking the ICC values into account, every static and
dynamic variable analysed showed moderate to excellent reliability and the CoV
values were all below 12%. CONCLUSION: Measurements of either static or dynamic
plantar pressure variables with the S-Plate(R) pressure platform show good
repeatability and reliability, and so it is useful for comparing steps within and
between sessions.
CI - Copyright (c) 2016 Elsevier B.V. All rights reserved.