Assessment of Hitachi HME-20 pulse and blood-pressure monitor
NORTHCOTE RJ; O'DONOGHUE J ; BALLANTYNE D
J MED ENG TECHNOL , 1985, vol. 9, n° 1, p. 10-13 Doc n°: 21462 Localisation : Documentation IRR Descripteurs : VD - MATERIEL - GENIE BIOMEDICAL, VA1 - PRESSION FORCE The performance of the Hitachi HME-20 pulse and blood-pressure (BP) monitor in comparison with direct intra-arterial BP recording and electrocardiographic monitoring is described. Highly significant (p < 0.001- correlations were found between intra-arterial systolic and diastolic pressures and pressures recorded by the Hitachi monitor. Similarly, the electrocardiographically computed heart-rate and that given by the Hitachi monitor were significantly correlated (p < 0.001). Systolic blood-pressure was underestimated by a mean of - 12 mmHg and tended to become more erroneous when intra-arterial pressure was > 150 mmHg. These results are comparable to more expensive pulse and blood-pressure monitors. We conclude that the instrument can reproduce a satisfactory estimate of heart-rate and blood-pressure and may be of particular use when a change of blood pressure is of prime importance, rather than an absolute measurement Langue : ANGLAIS Identifiant basis : 1985002334 |