Radioisotopic control for baclofen pump catheter failure

H
LE BRETON F; DAVIET JC; MONTEIL J
SPINAL CORD , 2001, vol. 39, n° 5, p. 283-285
Doc n°: 102750
Localisation : Centre de Réadaptation de Lay St Christophe , en ligne
Descripteurs : AE21 - ORIGINE TRAUMATIQUE Url : http://www.nature.com/sc/archive/index.html

Case report of Baclofen pump catheter failure investigated by radioisotope injection. OBJECTIVES: To report a safe and reliable method for evaluating catheter dysfunction. SETTING: France. METHODS: Single case report of failure of Baclofen pump investigated by radioisotope injection. RESULTS: The injection demonstrated the block in the catheter. The catheter failure was not visualised by plain X-ray nor by filling the pump with radio-opaque solution. CONCLUSION: Catheter failure is a common cause of intrathecal drug delivery problems and may be difficult to diagnose. When catheter disconnection, kink, or dislodgement is not visible on X-ray, radioisotopic control is a safe and reliable method for assessment.